A 12.5 GHz-Spaced Optical Frequency Comb Spanning >400 nm for near-Infrared Astronomical Spectrograph Calibration

Abstract
A 12.5 GHz-spaced optical frequency comb locked to a Global Positioning disciplined oscillator for near-IR spectrograph calibration is presented. The comb is generated via filtering a 250 MHz-spaced comb. Subsequency nonlinear broadening of the 12.5 GHz comb extends the wavelength range to cover 1380 nm to 1820 nm, providing complete coverage over the H-band transmission widow of Earth's atmosphere. Finite suppression of spurious sidemodes, optical linewidth and instability of the comb have been examined to estmiate potential wavelength biases in spectrograph calibration. Sidemode suppression varies between 20 db and 45 dB, and the optical linewidth is ~350 kHz at 1550 nm. The comb frequency uncertainty is bounded by +/- 30 kHz (corresponding to a radial velocity of +/- 5 cm/s), limited by the Global Positioning System disciplined oscillator reference. These results indicate this comb can readily support radial velocity measurements below 1 m/s in the near-IR.
0
AI overview
Reviews
Citation graph
Structured
This paper does not have an AI preview yet
Rating
No ratings yet
Rate
Community rating
—
/ 5
No ratings yet
Rating
—
Soundness
—
Presentation
—
Contribution
—
Confidence
—
Discussion
Top
Newest
Questions
Replication
Author
No comments yet — be the first