MCP
arXivarXiv
Preprint
Published: February 6, 2015
nucl-ex

PMT signal increase using a wavelength shifting paint

Abstract
We report a 1.65 times increase of the PMT signal and a simple procedure of application of a new wavelength shifting (WLS) paint for PMTs with non-UV-transparent windows. Samples of four different WLS paints, made from hydrocarbon polymers and organic fluors, were tested on a 5-inch PMT (ET 9390KB) using Cherenkov radiation produced in fused silica disks by 106^{106}Ru electrons on a `table-top' setup. The best performing paint was employed on two different types of 5-inch PMTs (ET 9390KB and XP4572B), installed in atmospheric pressure CO2_2 gas Cherenkov detectors, and tested using GeV electrons.