Electroluminescence pulse shape and electron diffusion in liquid argon measured in a dual-phase TPC

Abstract
We report the measurement of the longitudinal diffusion constant in liquid argon with the DarkSide-50 dual-phase time projection chamber. The measurement is performed at drift electric fields of 100 V/cm, 150 V/cm, and 200 V/cm using high statistics 39^{39}Ar decays from atmospheric argon. We derive an expression to describe the pulse shape of the electroluminescence signal (S2) in dual-phase TPCs. The derived S2 pulse shape is fit to events from the uppermost portion of the TPC in order to characterize the radial dependence of the signal. The results are provided as inputs to the measurement of the longitudinal diffusion constant DL, which we find to be (4.12 ±\pm 0.04) cm2^2/s for a selection of 140keV electron recoil events in 200V/cm drift field and 2.8kV/cm extraction field. To study the systematics of our measurement we examine datasets of varying event energy, field strength, and detector volume yielding a weighted average value for the diffusion constant of (4.09 ±\pm 0.09) cm2^2 /s. The measured longitudinal diffusion constant is observed to have an energy dependence, and within the studied energy range the result is systematically lower than other results in the literature.
0
AI overview
Reviews
Citation graph
Structured
This paper does not have an AI preview yet
Rating
No ratings yet
Rate
Community rating
—
/ 5
No ratings yet
Rating
—
Soundness
—
Presentation
—
Contribution
—
Confidence
—
Discussion
Top
Newest
Questions
Replication
Author
No comments yet — be the first