Targeting the Parkinsonian pathological gait scenario, this study proposes a CNN-BiLSTM-based deep learning framework that uses an optimized IMU sensor configuration to achieve high-accuracy estimation of bilateral vertical ground reaction forces, and completes an ablation validation covering 1023 sensor configurations.
Current traditional approaches to IMU-based ground reaction force estimation mainly comprise three technical routes: extended Kalman filtering, musculoskeletal modeling, and inverse dynamics models based on Newton-Euler equations. The common advantage of these methods is physical interpretability, but they generally rely on predefined kinematic and dynamic assumptions and have clear limitations in adapting to complex or non-periodic movements. These physics-based models also suffer from two typical defects: first, they struggle to simultaneously cover gait dynamics estimation across different walking speeds and elderly gait scenarios, requiring more complex models to adapt; second, pathological gait such as in Parkinson's disease violates multiple basic assumptions of physical models, including gait symmetry and steady-state periodicity, further limiting the effective deployment of these schemes.
Two clear research gaps remain in the field of IMU-based ground reaction force estimation for Parkinsonian pathological gait. First, the vast majority of existing studies have validated on healthy populations, and the evidence accumulated for pathological gait is very limited; Parkinson's disease patients exhibit stronger inter-individual heterogeneity and stride-to-stride fluctuation characteristics, greatly increasing the difficulty of accurately estimating ground reaction forces, and the robustness of existing IMU-based estimation schemes in this population has not been sufficiently verified. Second, both mainstream methodological paradigms have adaptation deficiencies: physics-based approaches are constrained by predefined biomechanical assumptions and encounter clear bottlenecks when estimating bilateral ground reaction forces of Parkinson's patients across different walking tasks, and deep learning research covering this scenario is also clearly insufficient.
This study proposes a CNN-BiLSTM-based sequence-to-sequence regression framework for synchronously estimating bilateral vertical ground reaction forces (vGRF) from wearable IMU sensor data. The framework is trained and evaluated using gait data from both healthy control subjects and Parkinson's subjects, covering two validation settings: within-subject and cross-subject. In addition to experiments on the main wearable IMU dataset, the study also conducted an external cross-dataset validation on another independent Parkinson's gait dataset; all subsequent baseline model comparisons and sensor configuration analyses were performed under the cross-subject setting of the Parkinson's cohort.
This CNN-BiLSTM hybrid neural network architecture was specifically designed for sequence estimation tasks involving both embedded spatial features, and has previously been applied in activity recognition, joint kinematics estimation, and ground reaction force estimation. The model implemented in this study consists of three core functional modules connected in series: the first is a CNN-based feature extraction module that converts the high-dimensional time-series data collected by raw IMUs into a lower-dimensional, more compact feature representation; the second is a bidirectional LSTM memory module that, through bidirectional processing of the input sequence, fully captures long-range dependencies in the gait time-series data; the third is a set of fully connected layers responsible for mapping the temporal features to the final bilateral vGRF estimation output. You can use the figure below to intuitively trace the complete data flow from raw IMU input to the final estimation output and the operational logic of each layer.
This study uses three standardized metrics to evaluate the model's bilateral vGRF estimation performance: the mean absolute error MAE normalized to body weight (%BW), the relative root mean square error rRMSE, and the coefficient of determination between measured and estimated values. In terms of computation, MAE and are first calculated separately for the left and right vGRF channels of each segmented sample, and averaged to obtain segment-level metrics; then all segment results are aggregated at the individual subject level, and finally summarized at the cohort level to obtain population statistics. Depending on the distribution characteristics of the data, results are presented as mean ± standard deviation or median [interquartile range].
To verify the performance of the proposed CNN-BiLSTM model, under the full 13-IMU cross-subject setting of the main Parkinson's disease dataset, the study compared it with 5 baseline models under unified experimental conditions, including TCN, MLP, and Transformer, as well as two ablation variants derived from the target architecture: a pure CNN model and a pure BiLSTM model. All baseline models were evaluated under exactly the same experimental configuration, ensuring a consistent basis for performance comparison.
After unifying the definitions of evaluation metrics, in the external cross-dataset validation step, since the external dataset was not equipped with a head IMU and could not be directly aligned with the main dataset's 13-IMU setting, both datasets were uniformly trimmed to the 12 non-head IMU configurations they share, enabling generalization capability verification under matched conditions. Under this 12-IMU setting, the model achieved an MAE of 4.08 ± 2.49% BW and an of 0.88 ± 0.11 on the main dataset, and an MAE of 7.45 ± 3.85% BW and an of 0.86 ± 0.17 on the external dataset; the subsequent generalization analysis is based on the results under this unified setting.
The vGRF estimation model proposed in this study demonstrates strong cross-subject generalization in the Parkinson's disease population, with a coefficient of determination R² of up to 0.93 for the healthy control group and a corresponding R² of 0.91 for Parkinson's disease patients. The sensor configuration significantly affects the final estimation accuracy, and the optimal sensor placement locations differ clearly between Parkinson's patients and healthy controls. You can use the figure immediately following to intuitively understand the trends in estimation accuracy and key performance boundaries corresponding to different sensor quantities and placement locations.
For single-IMU configurations in the Parkinson's disease patient population, the placement achieving the highest accuracy is the LD position, but the median and average R² of this configuration are both lower than those of the optimal configurations with more sensors, and its estimation reliability is clearly limited. Based on both median and average accuracy metrics, the overall best-performing configuration uses 4 IMUs; adding further sensors beyond this number does not bring substantial improvement in estimation performance, as accuracy enters a plateau region. Statistical significance comparisons between adjacent sensor configurations show no significant performance differences among the groups using 2 to 9 IMUs.
The study further completed a comprehensive ablation experiment covering 1023 IMU combinations, systematically evaluating the impact of sensor quantity and placement on estimation accuracy. The results show that estimation accuracy improves steadily as the number of sensors increases, until the configuration reaches 4 sensors, after which newly added sensors bring only marginal gains. Notably, a simplified dual-IMU configuration with sensors placed only at the LD and FH positions can achieve near-optimal performance, with a corresponding R² of 0.91±0.09, striking a practical balance between estimation accuracy, lightweight sensor systems, and wearing comfort.
This study verifies that deep learning can effectively capture the complex gait dynamics associated with Parkinson's disease, achieving high estimation accuracy with in both healthy controls and Parkinson's disease patients, as well as across different grouping strategies. Within-subject analysis results show comparable performance between Parkinson's patients and healthy controls, with both achieving an of 0.98, indicating that the CNN-BiLSTM model successfully captures the inherent variability of pathological gait and mitigates the group-dependent bias commonly seen in neurologically disease-related biomechanical estimation tasks.
In the baseline model comparison, the proposed CNN-BiLSTM architecture demonstrates a clear advantage over traditional models in capturing the complex spatiotemporal features of pathological gait. Although CNN-BiLSTM is not itself a novel architecture, the systematic comparison and ablation analysis completed in this study provide task-specific empirical support for adapting this architecture to bilateral vertical ground reaction force estimation from wearable IMU data in the Parkinson's disease scenario. Ablation results show that removing either the convolutional module or the recurrent module leads to a drop in estimation performance, indicating that both components contribute to the final model's effectiveness.
This study has clear boundaries of applicability. First, although an independent external cross-dataset validation was completed, the overall stock of public Parkinson's gait datasets remains limited, and the external validation dataset used virtual IMU signals generated from motion capture rather than recordings from real wearable IMUs, so it cannot fully cover the various signal variations encountered in actual deployment scenarios; further validation on larger-scale, more diverse datasets collected entirely from real wearable devices is still needed to establish broader generalization capability. Second, this study did not include clinical indicators of disease severity, whereas the gait characteristics of Parkinson's patients vary significantly with the severity of motor symptoms.