Action-grounded tissue affordance enables anticipatory auto-framing that lowers surgeon cognitive workload during laparoscopic surgery

Abstract
In laparoscopy, surgeon gaze tracks where the instruments will act; easing this demand through visual attention modeling requires dense labels of those interaction loci. These encode tacit knowledge: experts converge on consensus loci yet struggle to state the rules. Here we show that such labels can be recovered from completed actions in surgical videos, in which recorded instrument trajectories are converted into dense, continuous supervision. DiffeoAfford grounds tissue affordance by attaching instrument tips to the tissue and transporting them through deformation using diffeomorphism-constrained tracking, matching context-informed annotators' accuracy. Trained on these labels and never on gaze, a real-time model aligns with surgeon gaze more closely in space and time than does camera-assistant gaze. The framework also transfers across procedures: on hysterectomy videos, a separately trained predictor reaches 95.16% directional consistency with subsequent camera motion. In 12 paired cholecystectomies (24 procedures), the auto-framing application AffordView, which proactively centers predicted targets in view, lowered surgeon cognitive workload on converging subjective, physiological, and behavioral measures, including a reduced number of verbal instructions to the camera assistant. Deriving supervision from action rather than manual annotation offers a scalable route to anticipatory assistance.
0
AI overview
Reviews
Citation graph
References · 5
Structured
This preview uses paper v1; the current paper is v2
Generated from v1 · the paper is now at v2

Overview

Addressing the dense annotation challenge of visual assistance systems for laparoscopic surgery, this paper proposes an action-anchored tissue affordance framework, DiffeoAfford, which retrospectively generates affordance labels from completed surgical procedures with accuracy matching manual annotation. AffordView, a predictive auto-viewing system trained on this framework, was clinically validated to reduce surgeon cognitive load.

Scenario Needs and Practical Pain Points of Visual Assistance in Laparoscopic Surgery

Minimally invasive surgery fundamentally changes the surgeon's perceptual-motor environment: the dissociation of visual and motor axes, the loss of tactile input, and the restricted two-dimensional field of view provided by the endoscope—which induces a tunnel vision effect—severely impair the surgeon's peripheral spatial awareness.

Under these scenario characteristics, precise control of the field of view is a necessary cognitive aid for supporting surgeons in simultaneously operating the endoscope and surgical instruments.

Core Framework Contributions of DiffeoAfford and AffordView

For the laparoscopic surgery scenario, the paper's first core innovation is DiffeoAfford, an automated affordance annotation framework requiring no per-frame manual annotation. Rather than relying on manual frame-by-frame labeling to obtain affordance labels, the framework retrospectively recovers labels from completed surgical procedures. Through a diffeomorphic-constrained tracking mechanism, it propagates recorded instrument contact information across deforming tissue, ultimately producing affordance labels whose accuracy is comparable to those produced by human annotators with full surgical context.

An affordance prediction model trained on the soft labels generated by DiffeoAfford can perform real-time inference. Although the model was never trained on gaze-related data, its predictions align with intraoperative surgeon gaze in both temporal and spatial dimensions better than the framing performance of human camera assistants, and its predictions reach 95.16% directional agreement with subsequent camera motion.

The paper's second core innovation is AffordView, a predictive auto-viewing assistance system that actively places the predicted center of surgical interaction at the center of the frame, leveraging surgeons' natural observation tendencies to provide a stable visual reference. Unlike traditional reactive instrument tracking schemes or explicit gaze-linked systems, AffordView requires no additional calibration procedure and does not require surgeons to deliberately control their gaze. This intent-aware framework consistently keeps task-relevant information near predictable, high-priority coordinates, sparing operators the top-down executive effort needed to counteract the cognitive burden of the display's inherent spatial offsets.

Diffeomorphic-Constrained Affordance Annotation and Predictive Viewing Operation

DiffeoAfford models cross-frame correspondence with a diffeomorphic deformation field fitted jointly to a global transformation. Prior affordance anchoring schemes for natural scenes relied solely on global transformations; such methods capture only dominant camera motion, with insufficient degrees of freedom to cover the non-rigid distortions of compliant soft tissue. By introducing a diffeomorphic deformation field, the model gains additional deformation degrees of freedom while inherently possessing smoothness, invertibility, and topology preservation, structurally avoiding crossings and folding artifacts that violate biological tissue continuity.

DiffeoAfford's automated annotation pipeline first invokes SAM2 for surgical instrument segmentation and extracts instrument tip position and length parameters. For soft tissue deformation scenarios, it establishes dense pixel-wise correspondence across frames via diffeomorphic-constrained tissue tracking combining a global transformation with a local deformation field. Finally, it aggregates the instrument tip points on the query frame and propagates them to all other frames through the displacement field, completing automated generation of the affordance hotspot dataset. Together with the figure, you can fully understand all modules and the flow of logic from dataset generation to real-time hotspot prediction and AffordView auto-viewing.

Fig. 1: Overview of the proposed affordance-guided view control framework, covering the automated annotation pipeline, prediction model, and AffordView auto-viewing application

AffordView's auto-viewing module enforces explicit field-of-view boundary constraints: when the center of the predicted region of interest approaches the image boundary, the system restricts the cropped image dimensions, requiring cropped width w≥0.6Ww \geq 0.6W and height h≥0.6Hh \geq 0.6H, corresponding to a maximum zoom factor limited to 1.67x, thereby avoiding abnormal over-cropping and over-zooming. The dedicated prediction model deployed for clinical evaluation was trained using 520 valid surgical segments from the Calot's triangle region generated by DiffeoAfford.

Experimental Results from Multi-Dimensional Clinical Evaluation

The paper quantitatively evaluated the directional consistency of predictive viewing, measuring it at 95.16%, validating that the system's predicted viewpoint adjustment targets effectively anticipate the objects driving actual viewpoint adjustments. The study further compared the method with an instrument-tracking-based target prediction baseline, finding that instrument tracking only achieves comparable accuracy near the moment a motion event occurs; as the temporal gap between prediction and actual action grows, its prediction accuracy decays rapidly. The core reason is that before a camera adjustment occurs, the instrument tip is usually already very close to the target operating location.

In the intraoperative user experience evaluation, the proposed method showed statistically significant advantages across multiple dimensions including smoothness, stability, and plausibility of zooming logic. Surgeon preference for the proposed method was 75.4% on the smoothness dimension and 75.7% on the stability dimension, with all preferred-dimension statistical tests satisfying p<0.001. The method only failed to show a clear user preference advantage on the zooming dimension, a result that directly led the research team to integrate an instrument-driven zoom module in the subsequent intraoperative evaluation version.

The study also verified the system's reduction of surgeon cognitive load from both subjective and objective dimensions. On objective physiological measures, EEG analysis was conducted using the theta/alpha ratio as an established physiological marker of cognitive load; the experimental group showed a significant decrease in this measure compared to controls, and occipital alpha power in the later stages of surgery also fell below the control group's levels. In subjective cognitive load assessment, an exact paired permutation test showed the experimental group's overall SURG-TLX score was significantly lower than the control's, with the largest score decreases on the mental demand and physical demand subscales; surgeons also rated the system's usefulness significantly higher during the surgical phases when the system was active. Together with the figure that follows, you can directly compare the two groups' score distributions across SURG-TLX subscales and the trends in usefulness ratings across surgical phases.

Fig. 2: Subjective measures: a and b are radar charts of cognitive load assessed via SURG-TLX; c shows surgeons' intraoperative usefulness ratings across surgical phases

Method Applicability Boundaries and Summary of Key Conclusions

The 2D diffeomorphic approximation used by the method has clear applicability preconditions. Intraoperative soft tissue motion is inherently a three-dimensional deformation of the visible surface, yet the entire pipeline estimates only the projection of this 3D deformation onto the image plane; modeling that projection as a 2D diffeomorphism is a deliberate approximation based on surgical observation geometry and does not require support from a full 3D reconstruction. Surface displacement can be decomposed into a component parallel to the image plane and a component along the optical axis; only when the out-of-plane component along the optical axis is sufficiently small will there be no depth-order inversions—that is, self-occlusions—that break the planar model, keeping the approximation valid.

The validity of this approximation is ensured separately for each independent, continuous surgical action segment. Within the interval of a single operative action, both the laparoscope's viewpoint and the scene's depth ordering remain stable, and out-of-plane displacement stays within a small range; when segment boundaries arise where viewpoint or scene topology changes—such as large camera shifts, instrument-induced occlusions, or tissue dissection—the relevant footage is not included in the same deformation field computation but is instead handled separately by the temporal segmentation module.

For residual assumption violations that the segmentation mechanism cannot cover, such as transient tissue folding or independent motion of secondary instruments, the anomalies manifest as displacement outliers and can be directly removed via a Mahalanobis distance filtering step. This 2D diffeomorphic modeling design both matches the imaging geometry of laparoscopy and the regularities of surgical workflows, avoids the additional computational cost of full 3D reconstruction, and sidesteps the adaptation limitations imposed by static-scene assumptions.