MCP
arXivarXiv
Preprint
Published: August 4, 2026
cs.CV

Dual-domain U-Nets with embedded back projection operators for motion-resolved 4D CBCT reconstruction

Abstract
Four-dimensional cone beam CT (4D CBCT) is important for image-guided radiation therapy of thoracic cancers, but its use is limited by long scan times, causing high patient dose and motion/sparse-sampling artifacts. We propose a deep learning method for motion-resolved 4D CBCT reconstruction from conventional free-breathing scans, without a respiratory signal or explicit projection binning. Our CNN takes free-breathing 3D CBCT projections as input and predicts a static volume at maximum inhalation plus ten displacement vector fields (DVFs) spanning a breathing cycle. The network extends U-Net: the encoder acts on filtered projection stacks, the decoder acts in the volume domain, and skip connections are replaced with non-trainable back-projection functions at multiple resolutions to transfer features between domains. The model is trained on simulated CBCT scans and evaluated on 11 unseen simulated patients and 13 clinical free-breathing scans. Two additional models (60 s and 6 s scans) were evaluated by clinical experts on three and two scans, comparing single phases of our 4D reconstruction to reference 3D SART-TV images for tumor and esophagus visibility. Experts preferred our method for tumor visibility (59% vs. 36% no preference, 5% reference) and esophagus visibility (47% vs. 42%, 11%). On simulated data, image quality matched SART-TV (mean RMSE: -1.19 HU, PSNR: +0.09 dB, SSIM: -0.009) while enabling 4D reconstruction. On clinical scans, our method showed sharper dynamic structures (e.g., diaphragm) and fewer motion streak artifacts than traditional reconstruction. This non-patient-specific CNN predicts static volumes and full 4D respiratory motion models from a single free-breathing scan, without a respiratory surrogate or projection binning, reducing motion artifacts while adding motion-modeling capability.
0
AI overview
Reviews
Citation graph
References · 40
Structured
This preview uses generator v1; v2 is available

Overview

Targeting radiotherapy image guidance, this paper proposes a dual-domain U-Net deep learning architecture embedding a non-trainable backprojection operator, which can perform motion-resolved 4D reconstruction directly from projections of routine clinical free-breathing CBCT scans, without external respiratory surrogate signals or explicit projection binning.

Scenario Background and Research Positioning of 4D CBCT Reconstruction

This study addresses the scenario of 3D CBCT projection data acquired in routine clinical free-breathing scans. Against the gap of the strong dependence of existing reconstruction pipelines on respiratory signals and explicit projection binning, it explores a deep learning pathway that can complete 4D CBCT reconstruction without either of these two prerequisites.

Core Architecture of the Dual-Domain U-Net Embedding a Backprojection Operator

The network can infer temporal phase information directly from projection data, the model is not patient-specific, and training and evaluation can be done on a mid-range GPU. You can use the figure below to fully review the end-to-end design logic of this dual-domain network from projection input to multiple outputs.

Fig. 1: Overview of the CNN-based CBCT reconstruction method, taking CBCT projection input and ultimately outputting a static volume and cyclic DVFs representing patient motion within the respiratory cycle to produce a 4D image.

The core design of this dual-domain U-Net is embedding a non-trainable backprojection operator in the skip connections: at the corresponding levels of the U-Net, projection-domain feature maps are converted to the volume domain via the embedded non-trainable backprojection function; after instance normalization, the backprojected feature maps are concatenated with the corresponding features in the volume-domain decoder path. These backprojection functions are implemented as custom CUDA kernels; with the input projections filtered by Shepp-Logan, the overall logic resembles FDK reconstruction. To support end-to-end network training, the backpropagation of these layers is implemented as forward projection operations.

Multi-Scenario Evaluation Results and Method Limitations

On simulated test data, compared with the traditional SART-TV 3D reconstruction, the method maintains stable image quality metrics while achieving motion-resolved 4D reconstruction. For the anatomical structures most affected by respiratory motion, the method improves the average segmentation Dice scores over SART-TV, while the Dice score for the static spinal cord structure also remains stable. You can use the figure that follows to directly compare the differences in Dice score distributions between the two methods across six anatomical structures: right lung, left lung, liver, heart, esophagus, and spinal cord.

Fig. 2: Comparison of segmentation Dice scores between SART-TV and this method: segmentation performance improves for organs affected by respiratory motion, while the segmentation score for the static spinal cord remains stable.

In real clinical scans without ground-truth labels, the method achieves overall better image quality than the SART-TV and MKB reconstruction schemes, with motion artifacts greatly suppressed and clearer contours of motion-related structures, while lacking the sparse-sampling artifacts visible in MKB reconstruction; it can reduce noise levels while preserving respiratory motion information.

A total of 11 clinical experts participated in the evaluation, including 7 radiation oncologists and 4 medical physicists. Through side-by-side comparison of SART-TV 3D images with static end-expiration images of the method's 4D reconstruction, the experts cast preference votes on two dimensions: tumor visibility and esophagus visibility. The voting results showed that 59% of experts preferred the method's reconstruction for tumor visibility, 36% found no obvious difference between the two reconstructions, and only 5% preferred the reference 3D reconstruction; for esophagus visibility, 47% of experts preferred the method's reconstruction.

The method completes 4D CBCT reconstruction using only projection data, with no need to pre-record respiratory signals or perform explicit projection binning. After being trained on a population dataset, the model requires no patient-specific adaptation. It has just over 2 million parameters in total, and 4D reconstruction of 10 phases can be completed in under 4 seconds on a mid-range consumer GPU. A known limitation of the method is a certain blurring effect in the reconstructed results, which may obscure some high-frequency details—a common issue from training with pixel-wise loss functions.

Rating
No ratings yet
Rate
Community rating
—
/ 5
No ratings yet
Rating
—
Soundness
—
Presentation
—
Contribution
—
Confidence
—
Discussion
Top
Newest
Questions
Replication
Author
No comments yet — be the first