Initial radiometric calibration of the High-Resolution EUV Imager (
HRIEUV) of the Extreme Ultraviolet Imager (EUI) instrument onboard Solar Orbiter
The paper reports the ground-based radiometric calibration of the HRI_EUV telescope aboard Solar Orbiter, establishing its beginning-of-life performance. End-to-end and component-level measurements at two synchrotrons validated a semi-empirical model of the spectral response. The telescope achieves a peak efficiency above 1 e⁻·ph⁻¹, detector noise of ≈3 e⁻ rms, and pixel saturation above 120 ke⁻, confirming state-of-the-art signal-to-noise and spatial resolution for EUV solar imaging.