Microliter viscometry using a bright-field microscope:
η-DDM
The paper introduces \eta-DDM, a microliter viscometry method that uses a bright-field microscope. It applies Differential Dynamic Microscopy (DDM) to the Brownian motion of colloidal tracers, calculating the mean-squared displacement from the intermediate scattering function without particle tracking. The steady-shear viscosity is then inferred via standard microrheology and the Cox-Merz rule. Validated on Newtonian fluids, colloidal suspensions, and a viscoelastic polymer solution, the method requires only microliter sample volumes, offering a convenient alternative to classical rheometry for biological and soft matter systems.