Calibration of the NEXT-White detector using
83mKr decays
Using $^{83m}\mathrm{Kr}$ decays, the NEXT-White detector’s drift-electron lifetime and position-dependent energy variations are corrected via calibration maps. The resulting energy resolution at 41.5 keV is (4.553 ± 0.010 (stat.) ± 0.324 (sys.))% FWHM in the full chamber and (3.804 ± 0.013 (stat.) ± 0.112 (sys.))% FWHM in a fiducial volume. Scaling by $1/\sqrt{E}$ gives corresponding resolutions at the $Q_{\beta\beta}$ energy of xenon double beta decay (2458 keV) of (0.516 ± 0.0014 (stat.) ± 0.0421 (sys.))% FWHM and (0.4943 ± 0.0017 (stat.) ± 0.0146 (sys.))% FWHM, well within the 1% target.