Simple, compact, high-resolution monochromatic x-ray source for characterization of x-ray calorimeter arrays
M. A. LeuteneggerM. E. EckartS. J. MoseleyS. O. RohrbachJ. K. BlackM. P. ChiaoR. L. KelleyC. A. KilbourneF. S. Porter
M. A. Leutenegger
M. E. Eckart
S. J. Moseley
S. O. Rohrbach
J. K. Black
M. P. Chiao
R. L. Kelley
C. A. Kilbourne
F. S. Porter
+9
This paper presents a simple, compact monochromatic x-ray source built from channel cut crystals and a commercial x-ray tube. The device selects the Kα1 line of the anode material, producing a collimated beam with a line-shaped image. It achieves theoretical energy resolutions of 0.125, 0.197, and 0.086 eV FWHM at 5.4 and 8.0 keV, respectively, and an experimental upper limit of 0.7 eV FWHM at 5.4 keV. The source is portable and suitable for characterizing x-ray calorimeter arrays with reduced systematic uncertainty.

arXiv ·
· August 14, 2020