Vector Measurements Using All Optical Scalar Atomic Magnetometers
RZRui ZhangRahul MhaskarKen SmithEasswar BalasubramaniamMark Prouty
RZRui Zhang
Rahul Mhaskar
Ken Smith
Easswar Balasubramaniam
Mark Prouty
+5
An all-optical scalar atomic magnetometer measures the magnetic field magnitude from the AC component of the probe signal while also extracting the polar angle (between the field and probe beam) from the probe DC signal and from the light-shift-induced field due to frequency modulation. With a 100 ms measurement time, a commercial miniaturized sensor achieves better than 0.02° sensitivity at optimal orientation, and better than 1° accuracy over 0–180° when checked against an optical encoder. Azimuthal angle measurement is also demonstrated using two orthogonal sensors.

arXiv ·
· November 17, 2020