Assembly and Test of the Gas Pixel Detector for X-ray Polarimetry
H. LiH. FengF. MuleriR. BellazziniM. MinutiP. SoffittaA. BrezG. SpandreM. PincheraC. SgroL. BaldiniR. She +1
H. Li
H. Feng
F. Muleri
R. Bellazzini
M. Minuti
P. Soffitta
A. Brez
G. Spandre
M. Pinchera
C. Sgro
L. Baldini
R. She
+13
This paper presents the in-house assembly and test of a gas pixel detector (GPD) for X-ray polarimetry, designed for the XIPE mission. The detector, assembled at Tsinghua University, achieves a good electric field uniformity. Filled with pure dimethyl ether (DME) at 0.8 atm, the measured energy resolution is 18% at 6 keV, scaling inversely with the square root of energy. The modulation factor reaches ~0.6 above 6 keV, consistent with simulations. The residual modulation is 0.30% ± 0.15% at 6 keV, implying systematic error below 1% at 99% confidence. Position resolution is about 80 μm FWHM, meeting XIPE requirements.