Absorption of Scintillation Light in a 100 ℓ\ell Liquid Xenonγγ Ray Detector and Expected Detector Performance

Abstract
An 800L liquid xenon scintillation γ\gamma ray detector is being developed for the MEG experiment which will search for μ+→e+γ\mu^+\to\mathrm{e}^+\gamma decay at the Paul Scherrer Institut. Absorption of scintillation light of xenon by impurities might possibly limit the performance of such a detector. We used a 100L prototype with an active volume of 372x372x496 mm3^3 to study the scintillation light absorption. We have developed a method to evaluate the light absorption, separately from elastic scattering of light, by measuring cosmic rays and α\alpha sources. By using a suitable purification technique, an absorption length longer than 100 cm has been achieved. The effects of the light absorption on the energy resolution are estimated by Monte Carlo simulation.